Novus DC80L Manual de usuario

Busca en linea o descarga Manual de usuario para Sensores Novus DC80L. NOVUS DC80L User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 13
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
1
INDEX
1. INTRODUCTION ........................................................................................... 1
2. SAFETY INFORMATION .............................................................................. 2
3. MODEL DC80T ............................................................................................. 2
3.1. MODEL DC80T GENERAL SPECIFICATIONS ..................................... 3
3.2. MODEL DC80T FRONT PANEL ............................................................ 3
3.3. MODEL DC80T OPERATING INSTRUCTIONS .................................... 4
4. MODEL DC80R ............................................................................................. 5
4.1. MODEL DC80R GENERAL SPECIFICATIONS .................................... 6
4.2. MODEL DC80R FRONT PANEL ........................................................... 6
4.3. MODEL DC80R OPERATING INSTRUCTIONS ................................... 7
5. MODEL DC80L ............................................................................................. 8
5.1. MODEL DC80L GENERAL SPECIFICATIONS ..................................... 9
5.2. MODEL DC80L FRONT PANEL ............................................................ 9
5.3. PROTECTION ...................................................................................... 10
5.4. MODEL DC80L OPERATING INSTRUCTIONS .................................. 10
6. MAINTENANCE .......................................................................................... 13
1. INTRODUCTION
The Novus Calibrators series DC80 are portable instruments designed to meet the
user needs for measuring and simulating signals such as voltage, current,
resistance, thermocouples and RTDs. The available options depend on the
particular model. They are:
DC80T, for Thermocouples and mV
DC80R, for RTDs and resistance
DC80L, for linear mV, V and mA (provides internal power supply for
powering 2-wire transmitters)
Vista de pagina 0
1 2 3 4 5 6 ... 12 13

Indice de contenidos

Pagina 1

1 INDEX 1. INTRODUCTION ... 1 2. SAFETY INFORMATION ...

Pagina 2 - 3. MODEL DC80T

10 5.3. PROTECTION The MODEL DC80L calibrator uses two protection fuses, rated 125 mA / 250 V, internally connected in series with the current termin

Pagina 3

11 DC mV and V generation 1. Press the power switch to turn on the Calibrator. 2. Press the INPUT / OUTPUT mode key to select OUTPUT mode . 3.

Pagina 4

12 b) Measurements using calibrator’s internal power supply for powering the loop The calibrator provides an internal 24 V power supply for measuring

Pagina 5 - 4. MODEL DC80R

13 b) Simulating a 2-wireTransmitter (sink transmitter) 1. Press the power switch to turn on the Calibrator. 2. Press the INPUT / OUTPUT mode key

Pagina 6

2 2. SAFETY INFORMATION To avoid possible electric shock or personal injury: • Use the Calibrator only as specified in this manual, or the protection

Pagina 7 - RTD Simulation

3 3.1. MODEL DC80T GENERAL SPECIFICATIONS Maximum voltage allowed between any terminal and earth ground or between the two input terminals: ...

Pagina 8 - 5. MODEL DC80L

4 3.3. MODEL DC80T OPERATING INSTRUCTIONS In thermocouple temperature simulation the calibrator generates the voltage as an actual thermocouple, cons

Pagina 9

5 4. MODEL DC80R The MODEL DC80R RTD Process Calibrator measures and simulates resistance values of 7 distinct types of RTDs, either in °C or °F. Simu

Pagina 10

6 4.1. MODEL DC80R GENERAL SPECIFICATIONS Maximum voltage allowed between any terminal and earth ground or between the two input terminals: ...

Pagina 11 - DC mA measurement

7 4.3. MODEL DC80R OPERATING INSTRUCTIONS RTD measurement 1. Press the power switch to turn ON the calibrator. 2. Press the INPUT / OUTPUT key to

Pagina 12

8 5. MODEL DC80L The DC80L is the calibrator model for measuring and simulating V, mV and mA signals. The DC80L is capable of measuring and generating

Pagina 13 - 6. MAINTENANCE

9 5.1. MODEL DC80L GENERAL SPECIFICATIONS Maximum voltage allowed between any terminal and earth ground or between the two input terminals: ...